32
Dran-View 6 User Guide
COMTRADE
Import
Wizard Dialog
Page 1
Enterprise
Edition Only
Data Type This column defines the basic nature of the sampling rate. For
a waveform channel, select “Y-Waveform”, for a Trend (like
a RMS value), select “Y-Trend”. The remaining options are
“X-Time” which you should not use for any channel in a
COMTRADE file and “Y-Digital” which is used to denote
Digital status channels in a COMTRADE file. “Y-Digital”
should not be used for analog channels.
Group Use this setting to group parameters together into individual
measurement points. The number “1” should always be used
for single-point measurements.
Example: A file contains voltages from two measurement
points: AV1, BV1, CV1 and AV2, BV2, CV2. In this case,
you need to set Group number=1 for parameters
AV1,BV1,CV1 and Group number=2 for parameters
AV2,BV2,CV2. As a result of using more the one group,
there will be several sites created in the Dran-View 6
document, wherein each site will contain an individual 3-
phase system measurement.
Prefix Select “None” if the channel is not scaled. Select ‘k’ (kilo) if
the data has been divided by 1000 prior to being stored (other
Greek prefixes are also available).
Continued on the next page