5-5
CH 5/ Start Menu
Automatic Setup
procedure
(continued)
Action... Result...
STEP 5: Select your desired
probe type.
NOTE: Use the same probes for
the following: TR2500 for
TR2500A, TR2510 for TR2510A,
TR2520 for TR2520A.
If the current probe that you are
using is not on the list, select
Other. Then press Scale to
define the CT scale factor on
the current probe. Proceed to
Step 6.
•Press OK to accept probe
selection and monitor current.
•Press Cancel to retain previous
probe settings and return to
previous screen.
MARK104
STEP 6: You need to set the CT
scale factor when using a current
probe that is not on the list.
Pressing Scale will show the
numeric window to enter the CT
scale factor on the current probe.
Use the keypad to set the CT scale
factor. See sample scale factor
calculation on page 5-10.
•Press OK to accept the CT
scale factor value. The Current
Probe Selection screen will be
displayed.
•Press ... .... if you want to clear
and retype numeric values.
•Press x1000 to multiply the
value by 1000.
•Press Cancel to discard
changes and return to the
previous screen (screen shown
on Step 5).
MARK104b