Dranetz Xplorer Guide de l'utilisateur Page 234

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 290
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 233
C-6
Battery Pack Replacement
Battery Pack Replacement
Introduction
The PX5 contains an easily replaceable internal battery pack. See Appendix D for
ordering information.
WARNING
Replace with Dranetz
NiMH battery pack BP-PX5 only.
ADVERTENCIA
Reemplace con batería Dranetz NiMH BP-PX5 solamente.
AVERTISSEMENT
Remplacer par la batterie Dranetz NiMH BP-PX5 exclusivement.
WARNUNG
Nur mit Dran
etz NiMH BP-PX5 Batteriesatz auswechseln.
NOTE
During normal operation, the battery pack will be slightly warm to the touch.
Procedure
Follow these steps to replace the battery pack.
Continued on next pa
Step Action
1 Press the PX5 power button to off.
2 Turn off power to the circuit being measured.
3 Disconnect voltage and current probes from the circuit and the PX5s
rear panel.
4 On the bottom of the PX5, push the tab to release the battery cover. Refer
to the diagram shown on page C-7.
5 Remove the cover.
6 Pull up on removal strap and remove battery pack.
7 Insert the new pack into the compartment making sure to observe
polarity markings.
8 Replace the cover and press down until it latches closed.
9 Discard the old battery pack in accordance with Environmental
Protection Agency (EPA) regulations.
10 Press the PX5 power button to on.
Vue de la page 233
1 2 ... 229 230 231 232 233 234 235 236 237 238 239 ... 289 290

Commentaires sur ces manuels

Pas de commentaire