C-4
Battery Pack Replacement
Battery Pack Replacement
Introduction
The PowerVisa contains an easily replaceable internal battery pack. See Appendix D
for ordering information.
WARNING
Replace with Dranetz-BMI NiMH battery pack BP-PX5 only.
ADVERTENCIA
Reemplace con batería Dranetz-BMI NiMH BP-PX5 solamente.
AVERTISSEMENT
Remplacer par la batterie Dranetz-BMI NiMH BP-PX5 exclusivement.
WARNUNG
Nur mit Dranetz-BMI NiMH BP-PX5 Batteriesatz auswechseln.
NOTE
During normal operation, the battery pack will be slightly warm to the touch.
Procedure
Follow these steps to replace the battery pack.
Step Action
1 Press the PowerVisa power button to off.
2 Turn off power to the circuit being measured.
3 Disconnect voltage and current probes from the circuit and the
PowerVisa is rear panel.
4 On the bottom of the PowerVisa, push the tab to release the battery
cover. Refer to the diagram shown on page C-5.
5 Remove the cover.
6 Pull up on removal strap and remove battery pack.
7 Insert the new pack into the compartment making sure to observe
polarity markings.
8 Replace the cover and press down until it latches closed.
9 Discard the old battery pack in accordance with Environmental
Protection Agency (EPA) regulations.
10 Press the PowerVisa power button to on.