B-9
APPENDIX B/ Technical Specifications
Parameter Settings in Each Monitoring Mode
Setup
parameters
The following table lists the default settings for the threshold parameters in each
monitoring mode.
Continued on next page
Parameters
Standard
Power Quality
Long-term
Timed Recording
Continuous
Data Logging
EN50160
RMS Variation
High Limit 110% Off Off 110%
Low Limit 90% Off Off 90%
Very Low Limit 10% Off Off 1%
RMS Pre-trigger
(cycles)
6 0 1 6
RMS In-Out
Post-trigger
(cycles)
6 10 (50 Hz);
12 (60 Hz)
2 6
RMS Out-In
Post-trigger
(cycles)
6 10 (50 Hz);
12 (60 Hz)
2 6
Waveform
Capture Pre-
trigger (cycles)
2 0 1 2
Waveform
Capture Post-
trigger (cycles)
2 10 (50 Hz);
12 (60 Hz)
2 2
Transients
Absolute
Instantaneous
Peak
170% Off Off 170%
Cycle-by-cycle
Waveshape
Magnitude
10% Off Off 10%
Cycle-by-cycle
Waveshape
Duration (% of
cycle)
10% Off Off 10%
Waveform
Duration
7% Off Off 7%